Pall Acropak 500 Capsule Filter, 0.8um/0.2um, Sterile, Pall 12991
search
  • Pall Acropak 500 Capsule Filter, 0.8um/0.2um, Sterile, Pall 12991

Pall Acropak 500 Capsule Filter, 0.8um/0.2um, Sterile, Pall 12991

$155.00

Pall Acropak 500 Capsule Filter, 0.8um/0.2um filter membrane, sterilized by gamma radiation, stepped hose barb connectors, single pack of 1 capsule filter.

Pall 12991

Quantity

  Security policy

Your ordering history and payment details are secure with us!

  Delivery policy

Worldwide delivery is available to most countries.

  Return policy

Click here to read our Returns Policy

PALL AcroPak 500 - 12991 - Materials of Construction

  • Filter Media: Supor [hydrophilic polyethersulfone (PES)] membrane

  • Housing, Vent Plug, and Filter Support: Polypropylene

  • Sealing Technology: Thermal bonding

  • Filling Bell: Polycarbonate

Pore Size

  • 0.8um prefilter / 0.2um final filter

Effective Filtration AreaL 500cm2

Dimensions

  • Housing Diameter: 6.9 cm (2.7 in.)

  • Overall Length (Without Filling Bell):
       500 Series: 14.5 cm (5.7 in.)

Inlet/Outlet Connections

  • 500 and 1000 Series: 6.4 - 12.7 mm (1/4 - 1/2 in.) stepped hose barb with filling bell on outlet

Typical Hold-Up Volume (with 10 psi air purge)

  • 500 Series: 30 mL

Maximum Operating Temperature

  • 60 °C (140 °F) at 2.1 bar (210 kPa, 30 psi)

Maximum Operating Pressure

  • Continuous: 4.1 bar (410 kPa, 60 psi) at ambient temperature

Recommended Integrity Test Minimum Bubble Point – Water

  • 0.2 μm: 3.5 bar (350 kPa, 51 psi)

  • 0.45 μm: 1.7 bar (170 kPa, 24 psi)

  • 60% IPA/40% H2O (v:v)

  • 0.1 μm: 2.4 bar (240 kPa, 35 psi)

Typical Water Flow Rate

  • Lpm at 0.1 bar (10 kPa) [Lpm/psi]:

Bacterial Retention

  • 0.2 μm Capsules: Lot samples retain a minimum of 107 cfu/cm2 of B. diminuta per modified ASTM F838, current revision

Endotoxin Level

  • < 0.25 EU/mL using Limulus Amoebocyte Lysate (LAL) test

Biological Safety

  • Passes United States Pharmacopeia (USP) Biological Reactivity Test, In Vivo <88>

Sterilization

  • Sterilized by gamma irradiation

  • If desired, autoclave once only prior to use at 121 - 123 °C (250 - 253 °F) for a maximum of 20 min.


PALL 12991

12991
New
Comments (0)
No customer reviews for the moment.