PALL AcroPak 500 - 12991 - Materials of Construction
- Filter Media: Supor [hydrophilic polyethersulfone (PES)] membrane
- Housing, Vent Plug, and Filter Support: Polypropylene
- Sealing Technology: Thermal bonding
- Filling Bell: Polycarbonate
Pore Size
- 0.8um prefilter / 0.2um final filter
Effective Filtration AreaL 500cm2
Dimensions
- Housing Diameter: 6.9 cm (2.7 in.)
- Overall Length (Without Filling Bell):
500 Series: 14.5 cm (5.7 in.)
Inlet/Outlet Connections
- 500 and 1000 Series: 6.4 - 12.7 mm (1/4 - 1/2 in.) stepped hose barb with filling bell on outlet
Typical Hold-Up Volume (with 10 psi air purge)
Maximum Operating Temperature
- 60 °C (140 °F) at 2.1 bar (210 kPa, 30 psi)
Maximum Operating Pressure
- Continuous: 4.1 bar (410 kPa, 60 psi) at ambient temperature
Recommended Integrity Test Minimum Bubble Point – Water
- 0.2 μm: 3.5 bar (350 kPa, 51 psi)
- 0.45 μm: 1.7 bar (170 kPa, 24 psi)
- 60% IPA/40% H2O (v:v)
- 0.1 μm: 2.4 bar (240 kPa, 35 psi)
Typical Water Flow Rate
- Lpm at 0.1 bar (10 kPa) [Lpm/psi]:
Bacterial Retention
- 0.2 μm Capsules: Lot samples retain a minimum of 107 cfu/cm2 of B. diminuta per modified ASTM F838, current revision
Endotoxin Level
- < 0.25 EU/mL using Limulus Amoebocyte Lysate (LAL) test
Biological Safety
- Passes United States Pharmacopeia (USP) Biological Reactivity Test, In Vivo <88>
Sterilization
- Sterilized by gamma irradiation
- If desired, autoclave once only prior to use at 121 - 123 °C (250 - 253 °F) for a maximum of 20 min.
PALL 12991