Medicom Quattro Autoclave Sterilization Pouches, 7.5 x 13", pk 200
$30.00
Medicom SafeSeal Quattro Autoclave Sterilization Pouches, 7.5 x 13", pack of 200
Catalog No. 88030-4
Security policy
Your ordering history and payment details are secure with us!
Delivery policy
Worldwide delivery is available to most countries.
Return policy
Click here to read our Returns Policy
Medicom SafeSeal Quattro Autoclave Sterilization Pouches, 7.5 x 13", pack of 200, 88030-4
For use in steam autoclave sterilization and gamma sterilization only.
Material: Blue tinted film sealed to paper with 2 sets of lead-free indicator inks and a polyester-backed adhesive tape
Pouches are made with high quality porous medical grade Kraft paper and a laminated 2 ply thick blue tinted film made from a combination of polypropylene and polyethylene. These materials are selected to ensure proper sterilant penetration. Pouches and corresponding packaging is latex free.
Safe-Seal Quattro self-sealing pouches meet CDC guidelines for effective sterilization
Now with internal multi-variable indicators and external process indicators for steam sterilization (see 2nd photo)
Responds to 3 variables: time, temperature and saturated steam
Comply with CSA/ISO 11140-1:2005 standard for external indicators and internal indicators.
Comply with ISO 11607 standard for pouch performance
Pouches have a 5/8” wide adhesive tape seal and a 3/8” wide thermal single seal to ensure the pouch maintains sterility after sterilization.
Pouches are compatible with large number of commonly used disinfectants
Sterility: Pouches maintain sterility of the device for up to one year post sterilization.
Usage: For Single-use only
The following characteristics are routinely verified in accordance with ISO 11607-1:
a). Seal integrity - uniform seal width, no over sealed areas, leaks, channels, wrinkles, cracks, folds, tears, pinholes in the sealed area.
b). Seal Strength - uniform seal strength ensures an aseptic opening without delamination of tearing of the film
Now with internal multi-variable indicators and external process indicators for steam sterilization (see 2nd photo)
Responds to 3 variables: time, temperature and saturated steam
Comply with CSA/ISO 11140-1:2005 standard for external indicators and internal indicators.
Comply with ISO 11607 standard for pouch performance
Pouches have a 5/8” wide adhesive tape seal and a 3/8” wide thermal single seal to ensure the pouch maintains sterility after sterilization.
Pouches are compatible with large number of commonly used disinfectants
Sterility: Pouches maintain sterility of the device for up to one year post sterilization.
Usage: For Single-use only
The following characteristics are routinely verified in accordance with ISO 11607-1:
a). Seal integrity - uniform seal width, no over sealed areas, leaks, channels, wrinkles, cracks, folds, tears, pinholes in the sealed area.
b). Seal Strength - uniform seal strength ensures an aseptic opening without delamination of tearing of the film
PACK of 200
88030-4
New
Comments (0)
No customer reviews for the moment.